Alcohol Test Devices

CMI Intoxilyzer 5000

For more than 25 years, CMI’s Intoxilyzer® line of evidential infrared spectrometry breath alcohol instruments has evolved into the STANDARD for accuracy, reliability and courtroom evidence.

They’re backed by CMI’s factory service and support. Intoxilyzer® 5000s are the instrument of choice in the majority of state breath alcohol programs in the U.S. and in several areas of Canada.

That tradition continues with the Intoxilyzer® 5000EN, the newest in CMI’s Intoxilyzer® 5000 line of products.


Specifications

 

FUNCTIONAL
> Power Switch: Activates breath tester.
> Start Test Switch: Brings the unit out of standby and activates the test mode sequence.
> Software Mode Selection: Features such as data entry, printing, test mode, digits displayed, etc. can be changed by accessing the features setup menu using the keyboard.
> Digital Display: A 14-segment, 16-character, dot matrix (vacuum fluorescent) alphanumeric display which signals the operation of the unit,alerts the operator to required actions, and gives the alcohol test concentration in weight per volume.
> Audible Tones: Signal the completion of an operation, the presence of a malfunction, an incorrect operational procedure, or an unfulfilled test requirement.
Printed Test Record: Provides a printed record of the date, model and serial number of the instrument, the test results and time of the test on a multi-copy card or single/double ply roll paper.
> Breath Sampling: Instrument automatically senses end expiratory air (alveolar) using the technique of slope detection in conjunction with a minimum volume and minimum time requirement.
> Keyboard: Keyboard is used to set time and date, set software mode switches, initiate maintenance and diagnostic routines, and input test data in the test record.
> Calibration: The instrument is factory-calibrated and does not require periodic calibration adjustments as do fuel cell-equipped instruments.
> Communications: The instrument is equipped with an RS-232 port for external computer interface.
> Recirculation: The instrument is equipped with two solenoids which recirculates simulator vapor and extends the life of the simulator solution.
> Standby Mode: Standby mode reduces dust accumulation in the instrument and increases component life by shutting down non-vital functions during inactive periods.
> Remote Turn-On: Allows the operator, via a phone modem, to change the instrument from standby mode to testing mode for remote diagnostics, control checks or record transfer.
> Breath Volume Measurement: A flow sensor allows breath volume information to be calculated and printed during each test.>. Gas Calibration Ready: The instrument is equipped to use dry gas calibration verification medium as well as wet bath simulation.
> Barometric Pressure Measurement: If dry gas is used as a calibration verification medium, a barometric sensor is included to provide automatic correction for barometric pressure.


PERFORMANCE
> Range: 0.005 to 0.450 BrAC grams/210 liters
> Accuracy: Better than federal requirements, ±3% or ±.003 BrAC, whichever is greater
> Precision: Standard deviation of 0.003 BrAC or better
> Test Time: Typically less than one minute (excluding data entry)


ELECTRICAL
> Input Voltage: 117 volt AC ±10% or 230 VAC ±10%
> Input Current: 1.5 amps @117 volts AC
> Fusing and Filtering: 3 amp fuse for overload protection, passive filter to meet FCC specifications

ENVIRONMENTAL
> Operating Temperature: 68oF to 86oF (20oC to 30oC)
> Storage Temperature: 32oF to 140oF (0oC to 60oC)
> Humidity: 10% to 90%, non-condensing

OPTICAL
> Light Source: Tungsten filament in halogen gas enclosed by a clear quartz envelope. Life expectancy is more than 10,000 hours.
> Absorption Wavelength: Narrow passband IR filters are used to measure infrared absorption at specific wavelengths yielding reference, alcohol and interferent detection.
> Cooled Detector: Single stage, thermoelectrically-cooled lead selenide detector with an integral thermistor for temperature regulation. Life expectancy is more than seven years.
> Light Path: Path length is 11.4 inches (28.9 cm).


MECHANICAL
> Dimensions: 18.75” wide x 17.35” deep x 5.75” high (47.6 cm wide x 44.1 cm deep x 14.6 cm high). With organizer stand: 36.35” wide x 19.25” deep x 10.5” high (92.3 cm wide x 48.9 cm deep x 26.7 cm high).
> Weight: Approximately 30 lbs. (13.6 kg)
> Sample Chamber: 11.4” (28.9 cm) with a volume of 81.4cc
> Breath Tube: Instrument is provided with external, detachable, heated breath tube.


COMPUTER BASED

> A multiprocessor system employing a microprocessor controls the general operation of the instrument from information display to printer operation, keyboard interface, and electro-mechanical functions. A separate microcontroller is used for optical signal processing which increases the system’s signal handling.

> The microprocessor includes 56K EPROM (erasable programmable read only memory), 32K of battery-backed RAM and 8K of Scratch RAM (Random Access Memory). Additionally, the microcontroller has a separate 64K EPROM available and a separate 8K scratchpad RAM. Every aspect of operation, from displaying and printing of information to the basic electrical and mechanical functions, is micro-computer controlled.


WARRANTY
> One year, parts and labor.
> Two- and three-year are optional.


OPTIONAL
> Filters: The instrument can be equipped with three or five filters (for enhanced interferent detection).
> Keyboards: The instrument can be equipped with either a standard PC type keyboard or a sealed membrane type which can be built into the front, sloped panel of the instrument.
> External Printer: A Centronics-type printer port can be installed in the instrument which allows for printing of test results to an external printer.
> Simulator Temperature Monitor: This allows the instrument, through software and hardware, to monitor the simulator solution temperature.
> Heated Simulator Hose: Temperature-regulated simulator hoses eliminate the possibility of condensation in inlet and/or return tubing.
> Communications Software: CMI’s COBRA program allows test results to be uploaded to a PC for record-keeping and data management. COBRA also provides remote diagnostic and calibration verification capabilities. COBRA is a Windows® based application.
> Card Reader with External Keyboard: Allows for the use of magnetic strip encoded driver’s licenses and certification cards.
> Organizer Stand: Provides a keyboard tray, simulator holder (or dry gas standard), and storage for printer cards and mouthpieces.




Would you like more information?

If you would like to find out more about our Alcohol Breath Testers or any of our products, please let us know. Call us toll free on 888.464.3872 or use our online form and we will be in contact with you shortly.